The current first-floor Loeffler building EEG collection first started in March-May 2022. And is used for [[: | The current first-floor Loeffler building EEG collection first started in March-May 2022. And is used for [[: | ||
- | For setup, see [[: | + | * For setup, see [[: |
+ | * [[: | ||
+ | * [[: | ||
===== Battery ===== | ===== Battery ===== | ||
Line 24: | Line 27: | ||
* normal headphones would interfere with signal | * normal headphones would interfere with signal | ||
- | ===== Bucket Test ===== | ||
- | |||
- | TODO | ||
===== Timing/TTL ===== | ===== Timing/TTL ===== | ||
Line 32: | Line 32: | ||
* TTL sent over parallel port from presentation computer to control unit. see [[: | * TTL sent over parallel port from presentation computer to control unit. see [[: | ||
- | ===== Faraday Cage ===== | + | ===== Biosemi to LNCD Layout |
- | The hardware was originally in Western Psych (formally/ | ||
- | | + | ===== Notes from Cortech Solutions ===== |
- | < | + | From Smith 2025-03-25 (correspondence with NR) with emphasis added |
- | /Volumes/ | + | |
- | </code> | + | > Unplugging, turning off, and turning back on cannot fix the CM out of range problem. Handling the cables may affect whether a partially broken cable works or does not. **When you perform a bucket test, it is important to <wrap tip> |
+ | > When the blue light flashes with electrodes connected, | ||
- | ===== Biosemi to LNCD Layout ===== | ||
+ | > If a faulty electrode is connected to the system but not touching the body, the blue CM in Range light could be on if everything else was connected well. I am actually not certain whether a faulty EXG electrode being connected to the system but not the body would cause the CM in Range light to flicker. But, when that **faulty electrode is connected to the body, it would cause the CM in Range light to flicker if it was already burning solid blue**. If the CM in Range light was not already burning solid blue, then there could be a different reason, namely whether CMS and DRL were well connected to the body.. | ||