| + | ====== EEG at Loeffler ====== | ||
| + | |||
| + | The current first-floor Loeffler building EEG collection first started in March-May 2022. And is used for [[: | ||
| + | |||
| + | * For setup, see [[: | ||
| + | * [[: | ||
| + | * [[: | ||
| + | * [[: | ||
| + | * [[: | ||
| + | |||
| + | ===== Battery ===== | ||
| + | |||
| + | * ADC-Box Battery should be charged between sessions | ||
| + | * Make sure that the orange (charging) light is always on when plugged in | ||
| + | * Dr. Salisbury' | ||
| + | |||
| + | 2024-12-13 - battery loosing charge capacity? | ||
| + | |||
| + | {{: | ||
| + | |||
| + | ===== Audio ===== | ||
| + | |||
| + | * Tasks | ||
| + | * [[tools: | ||
| + | * [[tools: | ||
| + | * Hardware: Task PC -> 35mm aux -> RCA Adapter -> [[https:// | ||
| + | Alternative from [[https:// | ||
| + | * transducer/ | ||
| + | * normal headphones would interfere with signal | ||
| + | |||
| + | ===== Tasks ===== | ||
| + | |||
| + | As of 2025-10-09, [[: | ||
| + | * [[tools: | ||
| + | * [[tools: | ||
| + | * [[: | ||
| + | * [[: | ||
| + | * [[: | ||
| + | * [[: | ||
| + | ===== Timing/TTL ===== | ||
| + | |||
| + | * TTL sent over parallel port from presentation computer to control unit. see [[: | ||
| + | |||
| + | ===== EEG Cap Layouts ===== | ||
| + | ==== Layouts ==== | ||
| + | |||
| + | |||
| + | ^Front | ||
| + | |{{.: | ||
| + | |||
| + | ^Habit/7T cap ^ SPA cap ^ | ||
| + | |{{ : | ||
| + | |||
| + | |||
| + | |||
| + | ==== Habit ==== | ||
| + | * Standard BESA 64 channel layout | ||
| + | * 6 extra channels added to cap from Pete Bachman (added as external channels and are removed from preprocessing pipeline) | ||
| + | * Channels 65 and 66 are mastoids | ||
| + | * use '' | ||
| + | eeglabpath = fileparts(which(' | ||
| + | cap_location = fullfile(eeglabpath,'/ | ||
| + | </ | ||
| + | |||
| + | ==== 7T ==== | ||
| + | |||
| + | * Standard BESA 64 channel layout (some have 128 channels) | ||
| + | * 6 extra channels added to cap from Pete Bachman (added as external channels and are removed from preprocessing pipeline) | ||
| + | * Channels 65 and 66 are mastoids | ||
| + | * 64 channel layout: '' | ||
| + | eeglabpath = fileparts(which(' | ||
| + | cap_location = fullfile(eeglabpath,'/ | ||
| + | </ | ||
| + | * 128 channel layout: '' | ||
| + | correction_cap_location = hera(' | ||
| + | |||
| + | |||
| + | ==== SPA ==== | ||
| + | |||
| + | |||
| + | * Standard Biosemi 64 channel layout (different than Habit and 7T) | ||
| + | * Locations can be found from Biosemi website: [[https:// | ||
| + | * We did not get new channel labels so we are using the old channel labels from Habit/7T see MATLAB preprocessing code for fix | ||
| + | * Currently only using 61/64 channels on new SPA cap (9/23/2026) | ||
| + | |||
| + | |||
| + | |Skipping cap '' | ||
| + | |Channel '' | ||
| + | |||
| + | ===== Notes from Cortech Solutions ===== | ||
| + | |||
| + | From Smith 2025-03-25 (correspondence with NR) with emphasis added | ||
| + | |||
| + | > Unplugging, turning off, and turning back on cannot fix the CM out of range problem. Handling the cables may affect whether a partially broken cable works or does not. **When you perform a bucket test, it is important to <wrap tip> | ||
| + | > When the blue light flashes with electrodes connected, <wrap tip> | ||
| + | |||
| + | |||
| + | > If a faulty electrode is connected to the system but not touching the body, the blue CM in Range light could be on if everything else was connected well. I am actually not certain whether a faulty EXG electrode being connected to the system but not the body would cause the CM in Range light to flicker. But, when that **faulty electrode is connected to the body, it would cause the CM in Range light to flicker if it was already burning solid blue**. If the CM in Range light was not already burning solid blue, then there could be a different reason, namely whether CMS and DRL were well connected to the body.. | ||