+ | ====== EEG at Loeffler ====== | ||
+ | |||
+ | The current first-floor Loeffler building EEG collection first started in March-May 2022. And is used for [[: | ||
+ | |||
+ | * For setup, see [[: | ||
+ | * [[: | ||
+ | * [[: | ||
+ | |||
+ | ===== Battery ===== | ||
+ | |||
+ | * ADC-Box Battery should be charged between sessions | ||
+ | * Make sure that the orange (charging) light is always on when plugged in | ||
+ | * Dr. Salisbury' | ||
+ | |||
+ | 2024-12-13 - battery loosing charge capacity? | ||
+ | |||
+ | {{: | ||
+ | |||
+ | ===== Audio ===== | ||
+ | |||
+ | * [[tools: | ||
+ | * [[tools: | ||
+ | * Using " foam eartips" | ||
+ | Alternative from [[https:// | ||
+ | )) | ||
+ | * transducer/ | ||
+ | * normal headphones would interfere with signal | ||
+ | |||
+ | |||
+ | ===== Timing/TTL ===== | ||
+ | |||
+ | * TTL sent over parallel port from presentation computer to control unit. see [[: | ||
+ | |||
+ | ===== Biosemi to LNCD Layout ===== | ||
+ | |||
+ | |||
+ | ===== Notes from Cortech Solutions ===== | ||
+ | |||
+ | From Smith 2025-03-25 (correspondence with NR) with emphasis added | ||
+ | |||
+ | > Unplugging, turning off, and turning back on cannot fix the CM out of range problem. Handling the cables may affect whether a partially broken cable works or does not. **When you perform a bucket test, it is important to <wrap tip> | ||
+ | > When the blue light flashes with electrodes connected, <wrap tip> | ||
+ | |||
+ | |||
+ | > If a faulty electrode is connected to the system but not touching the body, the blue CM in Range light could be on if everything else was connected well. I am actually not certain whether a faulty EXG electrode being connected to the system but not the body would cause the CM in Range light to flicker. But, when that **faulty electrode is connected to the body, it would cause the CM in Range light to flicker if it was already burning solid blue**. If the CM in Range light was not already burning solid blue, then there could be a different reason, namely whether CMS and DRL were well connected to the body.. | ||