====== Prisma SSR EyeLink ======
* The EyeLink Tray Mounted Camera needs to be placed before the study.
* We need to use the "fancy" first-surfaced mirror on the head coil.
* This mirror needs extra care. It will scratch and smudge easily. It is difficult to clean.
Both of these are stored in the Prisma one scan room closet.
===== Hardware log =====
* 2025-08-26 - P2 mirror fell out! scuffed
===== EyeLink SetUp =====
Roughly
- Connect network cables
- Set IP (P1, no need P2)
- calibrate with ''control.exe''
- clean up (remove network cable, undo IP on P1)
Also see https://wiki.mrrc.pitt.edu/doku.php?id=hardware:eyetracking:eyelink:taskpc_setup
''Prisma 2''s EPrime Computer IP does not need to be adjusted! But P1 does.
==== Connection Setup ====
- Ensure blue Ethernet coord is connected from the EyeLink Host PC to "EPrime" Task computer
- Ensure EyeLink Host PC is turned on
- Ensure Monitor and Keyboard working
- For P1: Confirm EyeLink mouse and Keyboard are accessible
- For P2: Ensure monitor and keyboard are switched to show the EyeLink computer (option #2)
==== Prisma 1 IP Address Change ====
- Open the ''Manually_Set_IP.txt'' file on the task pc desktop (for IP address reference later)
- Adjust Windows IP address settings
- Click on the **control panel**
- click **View network status and task** under //Network and Internet// section
- click **change adapter** on the left sidebar
- right-click on **local area connection** and select **Properties**
- enter the same ''luna'' creds used for login
- select **Internet Protocol 4** from the connection items
- copy and paste address from the readme (both)
100.1.1.2 -- is ip address
100.1.1.1 -- is default gateway
- click **OK** and **No** (do not restart)
{{:tools:eyetracking:pasted:20240829-121113.png?200}}
{{:tools:eyetracking:pasted:20240829-121141.png?200}}
{{:tools:eyetracking:pasted:20240829-121214.png?200}} \\
{{:tools:eyetracking:pasted:20240829-121241.png?200}}
{{:tools:eyetracking:pasted:20240829-121302.png?200}}
==== Setup Camera display ====
- click the **control** program form the desktop (blue eye icon)
- switch the projector display switch to "task computer" if not already there
- enter an ID
- push **enter** once the screen has flash to gray to display the eye camera feed on the projector
- push the **right arrow key** to show the full feed instead of just the eye
- confirm camera is in focus and thresholds are reasonable
- if pupil thresholding cannot be established with value above 65, the IR is likely not hitting the eye.
==== Clean up ====
- **RESET THE IP ADDRESS BACK TO UPMC**
- follow the same control panel -> network properties -> adapter settings as above but use 10.48.88.???
- remove the ethernet coord from the task computer (back to disconnected like before)
- Gentle reminder to switch back eyetracking monitor and keyboard are switched back to show original screen (option #1)
- save the (hopefully) amazing ET data then LogOff of Luna on EPrime